Multichannel analyzer
A multichannel analyzer (MCA) is an instrument used in laboratory and field applications to analyze an input signal consisting of voltage pulses.[1] MCAs are used extensively in digitizing various spectroscopy measurements, especially those related to nuclear physics, including various types of spectroscopy (alpha-, beta-, and gamma spectroscopy).
Operation

A multichannel analyzer uses a fast ADC to record incoming pulses and stores information about pulses in one of two ways:[1]
Pulse-height analysis
In pulse-height analysis (PHA) mode, incoming pulses are characterized based on their amplitude (peak voltage). The output spectrum is a histogram of these pulses, where the height of each channel corresponds to the number pulses counted within a narrow range of amplitudes. The resolution of the output spectrum depends on the number of channels of the MCA, which is on the order of a few thousand for typical instruments.
In alpha-, beta-, and gamma spectroscopy, PHA is used to measure the energy distribution of particles emitted in nuclear decay.[2] Incoming particles are absorbed by a detector medium and excite voltage pulses whose amplitudes are proportional to their energy.[citation needed] After many pulses have been counted, the output spectrum shows the energy distribution of the radiation incident on the detector.
Multichannel scaling mode
In multichannel scaling (MCS) mode, the MCA records a pulse count-rate over time. Unlike PHA, MCS does not differentiate pulses of different amplitudes. Instead, the MCA records all measured counts in one channel for a set time interval (called the "dwell time"), then switches to the next channel to record the subsequent time interval, and so on.
The internal control voltage signal used to switch channels when the dwell time elapses is often available to the experimenter and can be used to trigger changes in the experimental setup.[1] In this arrangement, the MCA acts as an X–Y recorder, observing changes in the count rate as a function of the controlled experimental parameter. For example, a Geiger counter connected to an MCA in MCS mode could be used to record the amount of ionizing radiation emitted by a neutron generator at different voltages.
Output interface
Once a histogram has been recorded, the data is sent to a computer, displayed on a screen on the MCA, or (in older models) sent directly to a printer.
Modern MCAs typically interface with a computer via USB or Ethernet, but some older or specialty models use RS-232 or PCI.
Sound card MCA
A USB sound card can serve as a cheap, consumer off-the-shelf ADC, a technique pioneered by Marek Dolleiser. The data is sent to the computer as normal sound and stored in a WAV file. Specialized software processes the "sound" to perform pulse-height analysis and multichannel scaling, forming a complete MCA.[3]
Sound cards have high-resolution but low-speed (up to 192 kHz) ADC chips, allowing for reasonable gamma spectroscopy performance for a low-to-medium count rate.[4] The "sound card spectrometer" has been further refined in amateur and professional circles.[5][6]
See also
References
- ^ a b c "THE MULTICHANNEL ANALYZER" (PDF). Western University. Retrieved 27 March 2016.
- ^ "Gamma Spectroscopy with a Scintillation Detector and Multichannel Analyzer" (PDF). American University in Bulgaria. Retrieved 27 March 2016.
- ^ "Software Downloads". www.gammaspectacular.com.
Marek works at The University of Sydney, with third year physics students, and developed PRA as an educational tool for his students.
Motisan, Radu (November 29, 2010). "NaI Scintillation Probe and Gamma Spectroscopy – PocketMagic". www.pocketmagic.net. – Cited for early mention of Marek Dolleiser's PRA software. - ^ Ibrahim, Maslina Mohd; Yussup, Nolida; Lombigit, Lojius; Rahman, Nur Aira Abdul; Jaafar, Zainudin (2014). Development of multichannel analyzer using sound card ADC for nuclear spectroscopy system. International Nuclear Science, Technology & Engineering Conference 2013 (iNuSTEC2013). pp. 50–53. doi:10.1063/1.4866103.
- ^ Kasani, H.; Ashrafi, S.; Ghal-Eh, N. (July 2021). "High count-rate digital gamma-ray spectroscopy using a low-cost COTS digitizer system". Radiation Physics and Chemistry. 184 109438. Bibcode:2021RaPC..18409438K. doi:10.1016/j.radphyschem.2021.109438. S2CID 233696398.
- ^ Kim, Sangrok; Kim, Taeyoon; Yang, Hyungjin (1 June 2022). "Design of a Low-Resolution Gamma-ray Spectrometer for Monitoring Radioactive Levels of Wastewater". Applied Sciences. 12 (11): 5613. doi:10.3390/app12115613.
Content Disclaimer
Informasi ini disarikan dari Wikipedia dan disajikan kembali untuk tujuan edukasi. Konten tersedia di bawah lisensi CC BY-SA 3.0. Kami tidak bertanggung jawab atas ketidakakuratan data yang bersumber dari kontribusi publik tersebut.
- The information displayed on this website is sourced in part or in whole from Wikipedia and has been adapted for the purpose of restating it. We strive to provide accurate and relevant information, however:
- There is no guarantee of absolute accuracy. Wikipedia is an open, collaborative project that can be edited by anyone, so information is subject to change.
- It is not intended to constitute professional advice. The content displayed is for informational and educational purposes only. For important decisions (e.g., medical, legal, or financial), please consult a professional.
- Content copyright. Wikipedia is licensed under the Creative Commons Attribution-ShareAlike License (CC BY-SA). This means that content may be reused with appropriate attribution and shared under a similar license.
- Responsible use. Any risk arising from the use of information from this website is entirely the responsibility of the user.